FDA Recall Open, Classified

Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.

Recall: Z-0890-2026 · Initiated November 10, 2025

Recall

Recall Number
Z-0890-2026
Event Number
97971
Firm
Waldemar Link GmbH & Co. KG (Mfg Site) Oststr. 4-10 Norderstedt Germany
FEI Number
3003386935
Product Code
KRO
Status
Open, Classified
Root Cause
Process control
Initiated
November 10, 2025
Posted
December 9, 2025

Description

Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.

Reason

The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.

Action

On November 10, 2025, URGENT: MEDICAL DEVICE RECALL letters were emailed to distributors. Actions to be taken: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or [email protected] for any questions.

Distribution

US distribution to states of: AL, GA, and TX.

Quantity

18 units