FDA Recall Terminated

Customized Brius Appliances (custom metal orthodontic devices); BRIUS , RX Only, Non-Sterile, Custom Made Device, ID: 52345400UL

Recall: Z-0890-2022 · Initiated February 7, 2022

Recall

Recall Number
Z-0890-2022
Event Number
89569
Firm
Brius Technologies Inc.
FEI Number
3014680924
Product Code
DYJ
Status
Terminated
Root Cause
Device Design
Initiated
February 7, 2022
Terminated
October 18, 2024
Address
2611 Westgrove Dr, Carrollton, TX, 75006-2349

Description

Customized Brius Appliances (custom metal orthodontic devices); BRIUS , RX Only, Non-Sterile, Custom Made Device, ID: 52345400UL

Reason

Pontics, brackets, and customized orthodontic appliances were sold and distributed prior to completing the standard design control process, being documented, and released.

Action

On 2/28/22, the firm, Brius Technologies Inc., sent "URGENT: MEDICAL DEVICE RECALL" Letters (dated 2/21/2022) with an attached Acknowledgement and Receipt Response Form (Pontics, Patient Specific Brackets , and Customized Appliances), were sent to the respective Physicians via Fed-Ex first class trackable mail; a signed acknowledgement and receipt response was requested. Actions to be taken by the customer/user include the following: At this time, it is not recommended to remove the devices and interrupt the patient's treatment. However, physicians need to monitor their patients closely for adverse events. Additional Actions for Physicians and Patients with Pontics include in the letter: It is recommended that patients with Pontics be monitored for abnormal inflammation of tissues close to the location of the Pontics in the mouth. If the inflammation is thought to be an allergic response, it is recommended the removal of the device. Another cause of inflammation could be poor oral hygiene. In this case, it is not recommended to remove the Pontic. Additional Actions for Physicians and Patients with Patient Specific Brackets: All Brackets distributed prior to market release (7/22/21) have been removed from patients. No further actions are needed at this time. Additional Actions for Physicians and Patients with Customized Appliances: It is recommended that if the TAD (Temporary Anchorage Device) used with the customized device fails during treatment, the physician should remove the customized appliance and request a replacement appliance (that will not be customized to fit TADs). If customers have any questions they should contact: Elisabeth Miller [email protected],586-242-7718 Acknowledgment response forms should be completed and returned as follows: PLEASE EMAIL COMPLETED RESPONSE FORM TO: [email protected] OR MAIL TO: Attn: Michael Gregory 2611 Westgrove Dr. Ste109 Carrollton, TX 75006

Distribution

US Nationwide: CA, DC, MA, NJ, NY, TX

Quantity

13