FDA Recall Terminated

Intera Achieva 1.5T Nova Dual, Model Number 781173. Nuclear Magnetic Resonance Imaging System

Recall: Z-0889-2019 · Initiated December 21, 2018

Recall

Recall Number
Z-0889-2019
Event Number
82038
Firm
Philips Medical Systems Nederlands Veenpluis 4-6 Best Netherlands
FEI Number
3002807880
Product Code
LNH
Status
Terminated
Root Cause
Unknown/Undetermined by firm
Initiated
December 21, 2018
Posted
February 22, 2019
Terminated
November 19, 2020

Description

Intera Achieva 1.5T Nova Dual, Model Number 781173. Nuclear Magnetic Resonance Imaging System

Reason

There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.

Action

Urgent Medical Device Correction notification letters dated December 2018 were sent to customers.

Distribution

U.S. Nationwide distribution.

Quantity

11226 total