Batteries used with the LIFEPAK 20 defibrillator/monitor
Recall
- Recall Number
- Z-0887-2007
- Event Number
- 37692
- Firm
- Medtronic Emergency Response Systems, Inc.
- FEI Number
- 3015876
- Product Code
- MJK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 30, 2007
- Posted
- July 4, 2007
- Terminated
- May 4, 2011
- Address
- 11811 Willows Rd Ne, Redmond, WA, 98052-2003
Description
Batteries used with the LIFEPAK 20 defibrillator/monitor
Defibrillator internal battery cell may contain a short which could reduce battery capacity and cause the device to shut down without warning while operating on backup (DC) battery power.
Firm sent letters on 03/30/2007 dated 03/2007 via certified return receipt to consignees with affected LIFEPAK 20 devices. The firm advised consignees to use the Operator's Checklist in the Operating Instructions to perform daily user tests and function checks while operating the defibrillator on battery power. In addition consignees are advised to replace their battery every 1 to 2 years.
Distribution is worldwide. The device is designed to be used in healthcare provider facilities such as hospitals and clinic settings.
3002