SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824.
Recall: Z-0886-06
·
Initiated March 30, 2006
Recall
- Recall Number
- Z-0886-06
- Event Number
- 35258
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- LEH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 30, 2006
- Posted
- May 18, 2006
- Terminated
- February 29, 2012
- Address
- 200 S Kraemer Blvd, Brea, CA, 92822-6208
Description
SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824.
Reason
Beckman Coulter has received reports of patient samples and quality control recovery shifting low when using Vancomycin reagent lots M507219 and M511327. Internal QC recovery was as much as 15% lower than the assigned target value.
Action
A Product Corrective Action (PCA) letter was be sent the week of MAR 30, 2006 to all SYNCHRON Systems Vancomycin Reagent customers of lot numbers M507219 and M511327 informing them to discontinue use and discard reagent. (Letters were sent by US mail)
Distribution
Nationwide & Canada
Quantity
3147