SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824.

Recall: Z-0886-06 · Initiated March 30, 2006

Recall

Recall Number
Z-0886-06
Event Number
35258
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
LEH
Status
Terminated
Root Cause
Other
Initiated
March 30, 2006
Posted
May 18, 2006
Terminated
February 29, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824.

Reason

Beckman Coulter has received reports of patient samples and quality control recovery shifting low when using Vancomycin reagent lots M507219 and M511327. Internal QC recovery was as much as 15% lower than the assigned target value.

Action

A Product Corrective Action (PCA) letter was be sent the week of MAR 30, 2006 to all SYNCHRON Systems Vancomycin Reagent customers of lot numbers M507219 and M511327 informing them to discontinue use and discard reagent. (Letters were sent by US mail)

Distribution

Nationwide & Canada

Quantity

3147