EZ-DOP, diagnostic ultrasound imaging system
Recall: Z-0885-04
·
Initiated January 5, 2004
Recall
- Recall Number
- Z-0885-04
- Event Number
- 28472
- Firm
- DWL Systems, Inc.
- FEI Number
- 3003504574
- Product Code
- IYN
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 5, 2004
- Posted
- July 20, 2004
- Terminated
- September 10, 2004
- Address
- 100 Carpenter Drive, Suite 202, Sterling, VA, 20164
Description
EZ-DOP, diagnostic ultrasound imaging system
Reason
A certain sequence of data entries on the keypad could cause the ultrasound medical device to display incorrect transmission power.
Action
The firm notified consignees by mail on 1/5/2004 and advised of potential for inaccurate power display when the device is actually transmitting. Consignees were requested to return the device for a software upgrade.
Distribution
Units were distributed to end users worldwide including 67 domestic consignees.
Quantity
367 units