FDA Recall Terminated

EZ-DOP, diagnostic ultrasound imaging system

Recall: Z-0885-04 · Initiated January 5, 2004

Recall

Recall Number
Z-0885-04
Event Number
28472
Firm
DWL Systems, Inc.
FEI Number
3003504574
Product Code
IYN
Status
Terminated
Root Cause
Other
Initiated
January 5, 2004
Posted
July 20, 2004
Terminated
September 10, 2004
Address
100 Carpenter Drive, Suite 202, Sterling, VA, 20164

Description

EZ-DOP, diagnostic ultrasound imaging system

Reason

A certain sequence of data entries on the keypad could cause the ultrasound medical device to display incorrect transmission power.

Action

The firm notified consignees by mail on 1/5/2004 and advised of potential for inaccurate power display when the device is actually transmitting. Consignees were requested to return the device for a software upgrade.

Distribution

Units were distributed to end users worldwide including 67 domestic consignees.

Quantity

367 units