FDA Recall
Terminated
Monitor portion of the LifeVest device - specifically the Defibrillator printed circuit assembly (PCA) within the monitor. The LifeVest is indicated for adult patients who are at risk for sudden cardiac arrest and are not candidates for or refuse an implantable defibrillator.
Recall: Z-0883-2009
·
Initiated October 21, 2008
Recall
- Recall Number
- Z-0883-2009
- Event Number
- 50119
- Firm
- Zoll Lifecor Corporation
- FEI Number
- 3008642652
- Product Code
- MVK
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- October 21, 2008
- Posted
- January 28, 2009
- Terminated
- November 4, 2009
- Address
- 121 Freeport Rd, Pittsburgh, PA, 15238-3411
Description
Monitor portion of the LifeVest device - specifically the Defibrillator printed circuit assembly (PCA) within the monitor. The LifeVest is indicated for adult patients who are at risk for sudden cardiac arrest and are not candidates for or refuse an implantable defibrillator.
Reason
The charging circuit will not shut down properly.
Action
Physicians with active patients were notified of the product removal by letter via first-class mail dated 10/21/2008. The letter describes the problem and the potential risk. For additional information, contact ZOLL Lifecor Corporation at 1.800.543.3267.
Distribution
Nationwide. OUS: Germany, United Kingdom, Italy, and France.
Quantity
1685 units