FDA Recall Terminated

Minstrel Patient Lift; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model numbers HMB001-US and HMB002-US (with scale)

Recall: Z-0883-04 · Initiated January 30, 2003

Recall

Recall Number
Z-0883-04
Event Number
28866
Firm
Arjo, Inc.
FEI Number
1419652
Product Code
FSA
Status
Terminated
Root Cause
Other
Initiated
January 30, 2003
Posted
July 20, 2004
Terminated
May 12, 2004
Address
50 Gary Ave, Ste A, Roselle, IL, 60172-1684

Description

Minstrel Patient Lift; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model numbers HMB001-US and HMB002-US (with scale)

Reason

The tension pin was not installed in spreader bar pivot bolt to prevent the pivot bolt from migrating out of the spreader bar, allowing the spreader bar to detach from the lift.

Action

All affected end users were given a copy of the Customer Advisory Notice dated 1/30/03, during the Arjo salesmen visits to inspect the Minstrel lifts for the presence of the tension pin in the pivot bolt of the lift. The advisory informed the user of the incident where the spreader bar detached from the scale unit. The end user was requested to read the advisory and sign the bottom portion of it certifying that they read the advisory and that the Arjo salesman inspected their units for the presence of the tension pin, retaining a copy of the signed advisory notice and the inspeciton record.

Distribution

Nationwide. District of Columbia, New York, Washington, Iowa, Florida, North Carolina, Louisiana, Virginia, South Carolina, Minnesota and Colorado

Quantity

46 units