FDA Recall Terminated

Roche brand AccuData GTS System; Catalog number 404, material number 03136833001.

Recall: Z-0882-05 · Initiated April 18, 2005

Recall

Recall Number
Z-0882-05
Event Number
31619
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
CGA
Status
Terminated
Root Cause
Other
Initiated
April 18, 2005
Posted
June 11, 2005
Terminated
March 7, 2006
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Roche brand AccuData GTS System; Catalog number 404, material number 03136833001.

Reason

If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.

Action

Consignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.

Distribution

Nationwide and Guam.