FDA Recall
Terminated
Roche brand AccuData GTS Plus System; Catalog number 3000249, material number 03000249001.
Recall: Z-0881-05
·
Initiated April 18, 2005
Recall
- Recall Number
- Z-0881-05
- Event Number
- 31619
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- CGA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 18, 2005
- Posted
- June 11, 2005
- Terminated
- March 7, 2006
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
Roche brand AccuData GTS Plus System; Catalog number 3000249, material number 03000249001.
Reason
If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
Action
Consignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.
Distribution
Nationwide and Guam.