Stratus CS STAT Fluorometric Analyzer BHCG DilPak (CBHCG-D) in vitro diagnostic.
Recall
- Recall Number
- Z-0880-06
- Event Number
- 35112
- Firm
- Dade Behring Inc.
- FEI Number
- 2517506
- Product Code
- DHA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 3, 2006
- Posted
- May 19, 2006
- Terminated
- August 7, 2006
- Address
- 500 GBC Dr. PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101
Description
Stratus CS STAT Fluorometric Analyzer BHCG DilPak (CBHCG-D) in vitro diagnostic.
Foil sealing of the DilPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.
Customers were provided clear written instructions (04/06/2006) to discontinue use and discard remaining inventory of products over 2 months old. In addition, each customer is being contacted by phone as an effectiveness verification. Recalling firm is also instructing Stratus CS customers to process all samples and dilutions in duplicate.
Nationwide to hospitals and government accounts in addition to International accounts in Canada, Saudi Arabia, Bangladesh, UAE, India, Japan, Denmark, and Australia.
1151 cartons of 5 DilPaks