FDA Recall Terminated

Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm). Designed for cementless use in total hip arthroplasty.

Recall: Z-0878-2013 · Initiated January 15, 2010

Recall

Recall Number
Z-0878-2013
Event Number
64294
Firm
Medacta Usa Inc
FEI Number
3006639916
Product Code
MEH
Status
Terminated
Root Cause
Process control
Initiated
January 15, 2010
Posted
February 23, 2013
Terminated
February 23, 2013
Address
4725 Calle Quetzal, Ste B, Camarillo, CA, 93012-8429

Description

Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm). Designed for cementless use in total hip arthroplasty.

Reason

Medacta USA is recalling Versafitcup Impacting Ring because the impaction ring may not fit properly in the implant shell in all cases. This could result in surgical delay potentially increasing possible complications under anesthesia, having to ream up to a larger shell size leaving less bone than planned and repositiong/removal of the shell after initial impaction.

Action

Medacta USA sent notification letters dated 1/15/10 to Sales Representatives for customers that purchased the Versafitcup Impacting Ring. The letter informed the customers of the problem identified and the actions to be taken. Customers are informed that Medacta USA personnel will contact them by telephone to provide them with a RMA number. Customers are informed that they will receive a new set of redesigned/reworked impacting rings and a second near the end of the January 2010. Customers with questions are instructed to contact the firm at (805) 437-7085.

Distribution

Nationwide Distribution including the states of ID, CO, TX, WI, NY, CT, WA, CA, UT, and AZ.

Quantity

293 units