FDA Recall Open, Classified

8MM,MICRO BIPOLAR FORCEPS,IS4000 REF 471171

Recall: Z-0875-2025 · Initiated December 19, 2024

Recall

Recall Number
Z-0875-2025
Event Number
95938
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
NAY
Status
Open, Classified
Root Cause
Process change control
Initiated
December 19, 2024
Posted
January 10, 2025
Address
1266 Kifer Rd, Sunnyvale, CA, 94086-5304

Description

8MM,MICRO BIPOLAR FORCEPS,IS4000 REF 471171

Reason

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

Action

On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Distribution

U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and Puerto Rico. O.U.S: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mexico, Monaco, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Vietnam

Quantity

1,036