FDA Recall Terminated

Stratus CS STAT Fluorometric Analyzer BHCG TestPak (CBHCG) in vitro diagnostic.

Recall: Z-0874-06 · Initiated April 3, 2006

Recall

Recall Number
Z-0874-06
Event Number
35112
Firm
Dade Behring Inc.
FEI Number
2517506
Product Code
DHA
Status
Terminated
Root Cause
Other
Initiated
April 3, 2006
Posted
May 19, 2006
Terminated
August 7, 2006
Address
500 GBC Dr. PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Stratus CS STAT Fluorometric Analyzer BHCG TestPak (CBHCG) in vitro diagnostic.

Reason

Foil sealing of theTestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.

Action

Customers were provided clear written instructions (04/06/2006) to discontinue use and discard remaining inventory of products over 2 months old. In addition, each customer is being contacted by phone as an effectiveness verification. Recalling firm is also instructing Stratus CS customers to process all samples and dilutions in duplicate.

Distribution

Nationwide to hospitals and government accounts in addition to International accounts in Canada, Saudi Arabia, Bangladesh, UAE, India, Japan, Denmark, and Australia.

Quantity

126 cartons of 100 TestPaks