FDA Recall Terminated

OmniDiagnost System

Recall: Z-0874-03 · Initiated July 17, 2003

Recall

Recall Number
Z-0874-03
Event Number
26837
Firm
Philips Medical Systems Sales & Service Region No. America
FEI Number
1000524572
Status
Terminated
Root Cause
Other
Initiated
July 17, 2003
Posted
July 20, 2004
Terminated
January 12, 2005
Address
22100 Bothell Everett Highway, Bothell, WA, 98041

Description

OmniDiagnost System

Reason

Labeling of push button caps located on the control panel may not clearly identify the applied Source to Image distance (SID).

Action

On 7/17/03 the firm issued a letter dated June 12, 2003 to their customers advising that service personnel will visit the consignees and replace the buttons.

Distribution

The firm distributed devices to 81 hospitals and medical centers located throughout the U.S.

Quantity

52