FDA Recall Terminated

Light Source 300 Watt Xenon Lamp used in the following models: Viking EndoSite Xenon, Model 8050-1 and Boston Scientific Spyglass 300W Light Source (M00546190)

Recall: Z-0873-2007 · Initiated April 5, 2007

Recall

Recall Number
Z-0873-2007
Event Number
37793
Firm
Vision Systems Group, A Div of Viking Systems
FEI Number
1000305720
Product Code
KOG
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 5, 2007
Posted
June 6, 2007
Terminated
June 7, 2007
Address
134 Flanders Rd, Westborough, MA, 01581-1023

Description

Light Source 300 Watt Xenon Lamp used in the following models: Viking EndoSite Xenon, Model 8050-1 and Boston Scientific Spyglass 300W Light Source (M00546190)

Reason

Incorrect power entry module assembled into unit may cause burning to insulation, compromised electrical emissions and premature light source failure

Action

Viking notified 2 distributors and 1 end user by telephone on 3/15/07. Viking representatives performed on site visit to perform inspection and replace units if needed.

Distribution

Nationwide to CA and NC

Quantity

4