FDA Recall
Terminated
Elecsys Troponin T; catalog number 2017644
Recall: Z-0870-03
·
Initiated April 25, 2003
Recall
- Recall Number
- Z-0870-03
- Event Number
- 26091
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- MMI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 25, 2003
- Posted
- May 29, 2003
- Terminated
- June 19, 2003
- Address
- 9115 Hague Road, Indianapolis, IN, 46250-0457
Description
Elecsys Troponin T; catalog number 2017644
Reason
Hemoglobin interference limitation revised downward; hemolyzed samples may result in false negative results.
Action
An Urgent Product Correction letter was mailed to each account, during the week of 4/21/03, advising them not to use hemolyzed specimens with a hemoglobin level equal to, or greater than, 0.1 g/dL.
Distribution
United States