FDA Recall Terminated

Tecan Genesis RSP and NPS Instruments with barcode scanner type PosID2 option with firmware version 2.45 or lower.

Recall: Z-0869-03 · Initiated February 27, 2003

Recall

Recall Number
Z-0869-03
Event Number
25850
Firm
Tecan U S Inc
FEI Number
1037985
Product Code
JQW
Status
Terminated
Root Cause
Other
Initiated
February 27, 2003
Posted
May 29, 2003
Terminated
April 6, 2006
Address
4022 Stirrup Creek Rd, Ste310, Durham, NC, 27709

Description

Tecan Genesis RSP and NPS Instruments with barcode scanner type PosID2 option with firmware version 2.45 or lower.

Reason

Firmware defect. Under certain conditions a strip rack is no longer in the initial position after a motor overload of the barcode scanner PosID2.

Action

Tecan US notified consignees by registered letter on/about 3/13/2003. International accounts were notified via e-mail 3/13/2003 by Tecan Switzerland. Consignees were advised to contact the firm''s Technical Support line to coordinate the PosID firmware upgrade.

Distribution

Instruments requiring upgrade were distributed to various clinical laboratories and blood banks Nationwide and World wide .

Quantity

327 instruments require firmware upgrade.