FDA Recall Open, Classified

MODEL: X-RAY R/F SYSTEM FLUOROspeed X1

Recall: Z-0867-2022 · Initiated January 10, 2022

Recall

Recall Number
Z-0867-2022
Event Number
89596
Firm
Shimadzu Medical Systems
FEI Number
1000125362
Product Code
JAA
Status
Open, Classified
Root Cause
Software design
Initiated
January 10, 2022
Address
20101 S Vermont Ave, Torrance, CA, 90502-1328

Description

MODEL: X-RAY R/F SYSTEM FLUOROspeed X1

Reason

It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

Action

On or about 1/7/2022, the firm sent an "Urgent: Voluntary Medical Device Recall Notice" to customers informing them that there is a potential issue with the adjustment procedure of the generator on the SonialVision (G$ and safire) Systems and FluoroSpeed X1 Systems. As a result of the Recalling Firm's investigation and testing it was proven that there is a problem with the adjustment procedure documented for the calibration of the system xray generator. This problem occurs during the calibration of the fluoroscopic mode which could cause x-ray radiation dose rate to exceed regulation and/or standards. As a result of the investigation, it was determined that the calibration procedure for the generator to instruct the installer requires updating. This problem if not corrected could cause improper calibration which would lead to dose levels above regulations. Therefore, the Recalling Firm will release notification to update calibration instructions and to conduct calibration of the system following revised calibration procedure. Customers are instructed to: -Check their systems for the affected generators/systems. -Contact their distributor to schedule a system update and calibration. -Pay attention to the following points until the calibration/update is completed: If the SID (distance between the X-ray focus and the FPD input surface) is 1200 mm or more, it has been confirmed that the dose rate does not exceed the specified value. If SID 1200 mm cannot be used and SID 1100 mm is used, the following points must be kept in mind until readjustment. When fluoroscopy is performed under the condition where the fluoroscopy tube voltage is maximum, the dose rate may increase by up to 8 percent from the adjustment value at the time of installation and exceed the dose regulation and/or standard value. Therefore, pay attention to the fluoroscopy cumulative time so as not to perform fluoroscopy for a long time under the condition where the fluoroscopy tube voltage is

Distribution

U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A

Quantity

25 systems