FDA Recall
Terminated
VITROS VALP Reagent , REF 680 1710. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. Responsible firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, N.Y. 14626.
Recall: Z-0866-05
·
Initiated April 18, 2005
Recall
- Recall Number
- Z-0866-05
- Event Number
- 31942
- Firm
- Ortho-Clinical Diagnostics
- FEI Number
- 1000136573
- Product Code
- LEG
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 18, 2005
- Posted
- June 4, 2005
- Terminated
- October 13, 2005
- Address
- 100 Indigo Creek Dr, Rochester, NY, 14626-5101
Description
VITROS VALP Reagent , REF 680 1710. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. Responsible firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, N.Y. 14626.
Reason
Negatively biased results may be observed when the reagent is stored on the VITROS 5,1FS Chemistry System for less than seven days.
Action
Letters dated 4/18/2005 with instructions to process a mid-level QC fluid every eight hours when processing patient samples for valproic acid. Replacement product to be shipped as soon as available.
Distribution
37 U.S. consignees nationwide. Ortho-Clinical foreign affiliates in England, France and Spain.
Quantity
139 boxes