FDA Recall Terminated

VITROS VALP Reagent , REF 680 1710. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. Responsible firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, N.Y. 14626.

Recall: Z-0866-05 · Initiated April 18, 2005

Recall

Recall Number
Z-0866-05
Event Number
31942
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
LEG
Status
Terminated
Root Cause
Other
Initiated
April 18, 2005
Posted
June 4, 2005
Terminated
October 13, 2005
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS VALP Reagent , REF 680 1710. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. Responsible firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, N.Y. 14626.

Reason

Negatively biased results may be observed when the reagent is stored on the VITROS 5,1FS Chemistry System for less than seven days.

Action

Letters dated 4/18/2005 with instructions to process a mid-level QC fluid every eight hours when processing patient samples for valproic acid. Replacement product to be shipped as soon as available.

Distribution

37 U.S. consignees nationwide. Ortho-Clinical foreign affiliates in England, France and Spain.

Quantity

139 boxes