i-STAT HL7 Interface The HL7 Interface product is an accessory to the analyzer

Recall: Z-0865-2010 · Initiated September 29, 2009

Recall

Recall Number
Z-0865-2010
Event Number
53483
Firm
Abbott Point of Care Inc.
FEI Number
2245578
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
September 29, 2009
Posted
February 26, 2010
Terminated
July 1, 2010
Address
400 College Rd E, Princeton, NJ, 08540-6607

Description

i-STAT HL7 Interface The HL7 Interface product is an accessory to the analyzer

Reason

Abbott Point of Care has determined that it is possible under certain circumstances that some test records could have their status update messages mis-matched with another record, resulting in the incorrect patient name, order number of comment being included in the CDS test record only.

Action

Urgent Recall Notification letters were sent by Fed Ex to all US customers on September 29, 2009 and worldwide on October 1, 2009. Questions should be directed to Abbott Point of Care Technical Support at 1-800-366-8010, Option 1.

Distribution

Nationwide and Foreign: Germany, the United Kingdom, Switzerland and the Netherlands.

Quantity

36