i-STAT HL7 Interface The HL7 Interface product is an accessory to the analyzer
Recall: Z-0865-2010
·
Initiated September 29, 2009
Recall
- Recall Number
- Z-0865-2010
- Event Number
- 53483
- Firm
- Abbott Point of Care Inc.
- FEI Number
- 2245578
- Product Code
- JQP
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- September 29, 2009
- Posted
- February 26, 2010
- Terminated
- July 1, 2010
- Address
- 400 College Rd E, Princeton, NJ, 08540-6607
Description
i-STAT HL7 Interface The HL7 Interface product is an accessory to the analyzer
Reason
Abbott Point of Care has determined that it is possible under certain circumstances that some test records could have their status update messages mis-matched with another record, resulting in the incorrect patient name, order number of comment being included in the CDS test record only.
Action
Urgent Recall Notification letters were sent by Fed Ex to all US customers on September 29, 2009 and worldwide on October 1, 2009. Questions should be directed to Abbott Point of Care Technical Support at 1-800-366-8010, Option 1.
Distribution
Nationwide and Foreign: Germany, the United Kingdom, Switzerland and the Netherlands.
Quantity
36