Philips HDI 4000 Ultrasound System Part numbers 9896 053 15031 (8500-0066-01), 9896 053 18811 (8500-0080-01), 9896 053 22011 (8500-0087-01)
Recall: Z-0865-05
·
Initiated May 6, 2005
Recall
- Recall Number
- Z-0865-05
- Event Number
- 32064
- Firm
- Philips Ultrasound, Inc.
- FEI Number
- 3019216
- Product Code
- DPW
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 6, 2005
- Posted
- June 3, 2005
- Terminated
- May 21, 2008
- Address
- 22100 Bothell-Everett Hwy. P.O. Box 3003, Bothell, WA, 98041-3003
Description
Philips HDI 4000 Ultrasound System Part numbers 9896 053 15031 (8500-0066-01), 9896 053 18811 (8500-0080-01), 9896 053 22011 (8500-0087-01)
Reason
The acoustic display of Mechanical Index and Thermal Index is inaccurate - two transducers may exceed limits for acoustic intensity and MI.
Action
On May 6, 2005 the firm mailed a letter, via return receipt requested, to customers. The letter advised of the problem, and provided instructions to create new low power presets. The letter also advised that software would be issued on/before November 2005.
Distribution
The units were distributed worldwide. Domestic distribution included hospitals and medical centers.
Quantity
1225 units