FDA Recall Terminated

Philips HDI 4000 Ultrasound System Part numbers 9896 053 15031 (8500-0066-01), 9896 053 18811 (8500-0080-01), 9896 053 22011 (8500-0087-01)

Recall: Z-0865-05 · Initiated May 6, 2005

Recall

Recall Number
Z-0865-05
Event Number
32064
Firm
Philips Ultrasound, Inc.
FEI Number
3019216
Product Code
DPW
Status
Terminated
Root Cause
Other
Initiated
May 6, 2005
Posted
June 3, 2005
Terminated
May 21, 2008
Address
22100 Bothell-Everett Hwy. P.O. Box 3003, Bothell, WA, 98041-3003

Description

Philips HDI 4000 Ultrasound System Part numbers 9896 053 15031 (8500-0066-01), 9896 053 18811 (8500-0080-01), 9896 053 22011 (8500-0087-01)

Reason

The acoustic display of Mechanical Index and Thermal Index is inaccurate - two transducers may exceed limits for acoustic intensity and MI.

Action

On May 6, 2005 the firm mailed a letter, via return receipt requested, to customers. The letter advised of the problem, and provided instructions to create new low power presets. The letter also advised that software would be issued on/before November 2005.

Distribution

The units were distributed worldwide. Domestic distribution included hospitals and medical centers.

Quantity

1225 units