FDA Recall Terminated

Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN) and Uric Acid (URCA) methods on the Dimension Vista System

Recall: Z-0864-2010 · Initiated August 31, 2009

Recall

Recall Number
Z-0864-2010
Event Number
53323
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JIX
Status
Terminated
Root Cause
Other
Initiated
August 31, 2009
Posted
February 26, 2010
Terminated
December 13, 2010
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN) and Uric Acid (URCA) methods on the Dimension Vista System

Reason

Siemens Healthcare Diagnostics, Inc. has confirmed high accuracy shifts with Dimension Vista URCA Flex reagent cartridge (various lots) when calibrating with some vials from CHEM 1 calibrator lot 9DM001.

Action

An Urgent Field Safety Notice dated August 31, 2009, was issued to all affected customers. Customers were instructed to discontinue the use of Dimension Vista System CHEM 1 Cal lot 9DM001 and discard any remaining product. If customers have any question they can contact Technical Solutions at 1-800-241-0420.

Distribution

Nationwide and Foreign: Australia, Canada, Denmark, and Malaysia.

Quantity

1474 Cartons