FDA Recall Terminated

Beckman Coulter UniCel DxC Synchron Clinical System ISE Flow Cell. PN Numbers: A10405 (UniCel DxC 600); A11810 (UniCel DxC 600 PRO); A27318 (UniCel DxC 600i); A11816 (UniCel DxC 800); A11812 (UniCel DxC 800 PRO); A59102 (UniCel DxC 880i); A64903 (UniCel DxC 680i); A64871 (UniCel DxC 660i); A64871 (UniCel DxC 860i). Intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, or cerebrospinal fluid.

Recall: Z-0863-2010 · Initiated January 4, 2010

Recall

Recall Number
Z-0863-2010
Event Number
54344
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JJE
Status
Terminated
Root Cause
Device Design
Initiated
January 4, 2010
Posted
March 4, 2010
Terminated
April 17, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92821-6208

Description

Beckman Coulter UniCel DxC Synchron Clinical System ISE Flow Cell. PN Numbers: A10405 (UniCel DxC 600); A11810 (UniCel DxC 600 PRO); A27318 (UniCel DxC 600i); A11816 (UniCel DxC 800); A11812 (UniCel DxC 800 PRO); A59102 (UniCel DxC 880i); A64903 (UniCel DxC 680i); A64871 (UniCel DxC 660i); A64871 (UniCel DxC 860i). Intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, or cerebrospinal fluid.

Reason

Excessive Buildup of protein, bacteria and sample tube additives in the ISE flow cell may cause erroneous NA (sodium) results.

Action

An Urgent: Product Corrective Action letter dated January 4, 2010, was mailed to customers advising of erroneous sodium results derived from build up of protein, bacteria and additives in the ISE flow cell. The letter discussed the impact and actions to be taken along with some additional recommendations. Customers are advised to follow the new maintenance instructions included with the letter. The information in the letter should be shared with laboratory staff and the notification should be retained as part of the Quality System documentation. A copy of the letter should be provided to any other laboratory that affected product was forwarded to. The enclosed response form should be completed and returned. Questions should be directed to the Customer Support Center at 1-800-854-3633.

Distribution

Worldwide Distribution -- USA and Canada.

Quantity

2630