Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated Rochester, NY
Recall
- Recall Number
- Z-0862-2015
- Event Number
- 69717
- Firm
- Bausch & Lomb Inc
- FEI Number
- 1920664
- Product Code
- HQF
- Status
- Terminated
- Root Cause
- Software Manufacturing/Software Deployment
- Initiated
- November 4, 2014
- Posted
- December 23, 2014
- Terminated
- March 23, 2017
- Address
- 3365 Tree Court Industrial Blvd, Saint Louis, MO, 63122-6615
Description
Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated Rochester, NY
A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.
Bausch & Lomb contacted customers by phone on November 4, 2014. Customers were informed of the conditions required for the device to malfunction. Customers were also notified that a software update to address the situation will be installed in their system within the next few weeks, and that the firm's service team will be scheduling a visit to each facility to install this software update. Customers were also told that in the meantime they may continue to use the device. For further questions please call (585) 338-6839.
Worldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates.