Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades.
Recall: Z-0862-2009
·
Initiated November 17, 2008
Recall
- Recall Number
- Z-0862-2009
- Event Number
- 50499
- Firm
- Miltex, Inc.
- FEI Number
- 2518040
- Product Code
- GES
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 17, 2008
- Posted
- February 20, 2009
- Terminated
- February 23, 2009
- Address
- 589 Davies Drive, York, PA, 17402
Description
Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades.
Reason
Blade may break: Over milling of the blade resulted in tip more narrow than specified.
Action
An Urgent Medical Device Recall letter dated 11/17/08 with an accompanying fax back form was sent to each customer. Each distributor was instructed to review their inventory, complete the form, and return any affected product. The distributors were also instructed to conduct a sub recall notification.
Distribution
Nationwide Distribution and Canada.
Quantity
200 boxes