CombiSet True Flow Series, Hemodialysis Blood Tubing with Priming Set and Transducer Protectors, with Access Flow Reversing Connector (Twister) Product Code: 03-2994-6 24 sets / case CombiSet Bloodlines Including split-septum injection site, viral-retentive transducer protectors, & priming set & Twister device
Recall
- Recall Number
- Z-0859-2009
- Event Number
- 50491
- Firm
- Fresenius Medical Care North America
- FEI Number
- 3001451489
- Product Code
- KOC
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 25, 2008
- Posted
- January 23, 2009
- Terminated
- September 24, 2009
- Address
- 920 Winter St, Waltham, MA, 02451-1521
Description
CombiSet True Flow Series, Hemodialysis Blood Tubing with Priming Set and Transducer Protectors, with Access Flow Reversing Connector (Twister) Product Code: 03-2994-6 24 sets / case CombiSet Bloodlines Including split-septum injection site, viral-retentive transducer protectors, & priming set & Twister device
Port may crack and separate resulting in potential blood exposure/blood loss for the patient
Fresenius Medical contacted customers by e-mail,telephone and hard copy letter advising users to immediately discontinue use and return inventory. All internal customers were emailed on 11/25/08. All external customers were US Certified Mailed on 11/26/08. All internal customers were US Certified Mailed on 12/2/08 & 12/3/08.
Worldwide Distribution --- including USA and country of Canada.
57 units