FDA Recall Terminated

GE Signa Advantage nuclear magnetic resonance imaging system.

Recall: Z-0858-06 · Initiated August 19, 2005

Recall

Recall Number
Z-0858-06
Event Number
33320
Firm
GE Medical Systems, LLC
FEI Number
2126677
Product Code
LNH
Status
Terminated
Root Cause
Other
Initiated
August 19, 2005
Posted
May 10, 2006
Terminated
February 3, 2007
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Signa Advantage nuclear magnetic resonance imaging system.

Reason

During use of GE Signa nuclear magnetic resonance imaging system, a flexible Teflon tube that was part of the magnet vent system ruptured and allowed the magnet to vent into the scan room. Because of helium pressure inside the room, the scan room door could not be reopened after the patient was removed from the scan room.

Action

GE Healthcare is issuing an Urgent Safety Notification to all affected customer sites beginning 09/01/05. The notification describes the concurrent condition that must exist in order for the event to occur. Users are warned to be present at all times when scanning patients and to take emergency measures to vent the scan room should the magent quench. Beginning September 2005, all affected customers'' devices will have the Teflon venting tube replaced with a nonflexible venting tube.

Distribution

Nationwide and to countries OUS including: United Arab Emirates, Argentina, Australia, Brazil, Canada, China, Germany, Denmark, Egypt, Spain, France, United Kingdom, Ireland, Italy, Jordan, Japan, Korea, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Saudi Arabia, Sweden, Turkey and Tiawan

Quantity

644 (516 US, 128 OUS)