FDA Recall Terminated

Infant Nasal CPAP Cannula, Size 0; an Rx infant nasal prong continuous positive airway pressure (CPAP) cannula one inspiratory and one expiratory elbow connector; 10 units per case; Teleflex Medical, Research Triangle Park, NC 27709 U.S.A., Made in Mexico; The cannulas are sold as components in the following configurations: a) catalog No. 1690 and 41690 (International): Infant CPAP Set intended for use with mechanical ventilators, which includes one cannula with one inspiratory and one expiratory elbow connector, two 10mm to 7.5mm adapters, one knit cap and two 6 inch sections of Velcro securing tape; b) catalog No. 1683: Infant CPAP System intended for use with a water seal or spring loaded valve, and includes one cannula with one inspiratory and one expiratory elbow conenctor, two 4 ft. lengths of 10mm I.D. corrugated tubing, one 4 ft. pressure monitoring line, one 22mm to 10mm humidifier adapter, one knit cap and two 6 inch sections of Velcro securing tape;

Recall: Z-0857-2007 · Initiated March 23, 2007

Recall

Recall Number
Z-0857-2007
Event Number
37665
Firm
Teleflex Medical
FEI Number
3004959265
Product Code
BZD
Status
Terminated
Root Cause
Other
Initiated
March 23, 2007
Posted
June 7, 2007
Terminated
November 16, 2009
Address
2345 Waukegan Rd, Ste 120, Bannockburn, IL, 60015-1580

Description

Infant Nasal CPAP Cannula, Size 0; an Rx infant nasal prong continuous positive airway pressure (CPAP) cannula one inspiratory and one expiratory elbow connector; 10 units per case; Teleflex Medical, Research Triangle Park, NC 27709 U.S.A., Made in Mexico; The cannulas are sold as components in the following configurations: a) catalog No. 1690 and 41690 (International): Infant CPAP Set intended for use with mechanical ventilators, which includes one cannula with one inspiratory and one expiratory elbow connector, two 10mm to 7.5mm adapters, one knit cap and two 6 inch sections of Velcro securing tape; b) catalog No. 1683: Infant CPAP System intended for use with a water seal or spring loaded valve, and includes one cannula with one inspiratory and one expiratory elbow conenctor, two 4 ft. lengths of 10mm I.D. corrugated tubing, one 4 ft. pressure monitoring line, one 22mm to 10mm humidifier adapter, one knit cap and two 6 inch sections of Velcro securing tape;

Reason

The luer-port in the expiratory elbow may be occluded, which interferes with a ventilator's ability to detect an accurate pressure reading.

Action

Teleflex Medical sent Urgent Medical Device Recall letters dated 3/23/07 via UPS 2nd Day Air to the direct accounts, informing them that the luer-port in the expiratory elbow may be occluded, which interferes with a ventilator's ability to detect an accurate pressure reading. The accounts were requested to cease use of the affected units, remove them from use/stock, quarantine and destroy them. If the accounts further distributed the product, they were requested to conduct a sub-recall of the affected units from their customers. The accounts were instructed to complete and return by fax the enclosed Acknowledgement & Stock Status Form, indicating the number of units destroyed and if they are requesting replacement product. Any questions were directed to Linda Todd at 1-800-334-9751, ext. 4951.

Distribution

Worldwide, including USA, Puerto Rico, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, El Salvador, Germany, Guatemala, Hong Kong, India, Israel, Kuwait, Malaysia, Mexico, New Zealand, Paraguay, Peru, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, and Venezuela

Quantity

13,530 for all products