FDA Recall Terminated

Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU)

Recall: Z-0857-06 · Initiated November 16, 2005

Recall

Recall Number
Z-0857-06
Event Number
34050
Firm
GE Healthcare
FEI Number
2183066
Product Code
BSZ
Status
Terminated
Root Cause
Other
Initiated
November 16, 2005
Posted
May 9, 2006
Terminated
January 23, 2007
Address
3030 Ohmeda Drive PO Box 7550, zip 53707-7550, Madison, WI, 53718

Description

Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU)

Reason

Alarm sound-It is possible that the audible alarms from the speaker from the Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU) machine may dampen or disappear as a result of normal handling of the anesthesia machine. Visual alarms on the ADU are unaffected by this condition. The dampening or absence of audible alarms on the ADU may result in a delay in treatment.

Action

Customers were sent a letter dated November 16, 2005, informing them of the possible alarm failure and giving them short term and long term solutions. Short term, clinicians are to: 1) check the audio alarm prior to use; and 2) use the ADU in conjunction with functioning anesthetic agent monitor. In the long term a GE Healthcare representative will contact the facility and schedule a visit to implement the corrective action at no charge.

Distribution

Worldwide---USA-States of AZ, CA, IA, MI, TX, and WI ----International- Australia, Austria, Belgium, Finland, Germany, Israel, Italy, Jordan, Malaysia, Netherlands, Norway, Portugal, Slovakia, South Africa, and Turkey.

Quantity

113---(USA-61--International-52)