R2 Adult R2 Non-Radiolucent Multifunction Electrodes, Catalog/REF No. 3200-1715, NON-STERILE, MADE IN USA, CONMED CORPORATION, 525 FRENCH ROAD, UTICA, NY 13507 This device, when connected to a monitor/defibrillator has the ability to monitor the electro-cardio function (ECG) of the patient. The device, when electrically activated by the monitor/defibrillator can pace the heart rate of the patient and may also transmit the electrical current to the patient for cardio-conversion. This is a non-sterile, disposable device.
Recall
- Recall Number
- Z-0856-2012
- Event Number
- 60701
- Firm
- ConMed Corporation
- FEI Number
- 3001451571
- Product Code
- MLN
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- March 31, 2011
- Posted
- January 25, 2012
- Terminated
- January 26, 2012
- Address
- 525 French Road, Utica, NY, 13502
Description
R2 Adult R2 Non-Radiolucent Multifunction Electrodes, Catalog/REF No. 3200-1715, NON-STERILE, MADE IN USA, CONMED CORPORATION, 525 FRENCH ROAD, UTICA, NY 13507 This device, when connected to a monitor/defibrillator has the ability to monitor the electro-cardio function (ECG) of the patient. The device, when electrically activated by the monitor/defibrillator can pace the heart rate of the patient and may also transmit the electrical current to the patient for cardio-conversion. This is a non-sterile, disposable device.
The specific lot was assembled without the rivet cover.
ConMed Corporation sent a Urgent Medical Device Recall Letter dated March 25, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to immediately stop use of the device and return it to ConMed Corporation. Please review your inventory for any of the devices listed on Attachment I. If you do not have any devices to return, please follow the instructions in the letter and complete Attachment II, indicating you have no devices and fax it to 315-624-3225. Please complete the attached Business reply form. We apologize for any inconvenience this will cause you or your staff. For further questions please call (315) 797-8375.
Worldwide Distribution -- USA (nationwide) and the countries of Canada, Belgium, Germany and Hong Kong.
850 sets (2 pads per set in individual pouch)