FDA Recall Terminated

Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667

Recall: Z-0856-06 · Initiated March 16, 2006

Recall

Recall Number
Z-0856-06
Event Number
34900
Firm
B & K Medical Systems, Inc.
FEI Number
1000305740
Product Code
IYN
Status
Terminated
Root Cause
Other
Initiated
March 16, 2006
Posted
May 9, 2006
Terminated
October 12, 2006
Address
250 Andover St, Wilmington, MA, 01887-1048

Description

Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667

Reason

Due to a problem with calibration, the device can produce up to 50% elevated acoustic output and up to 2 degrees higher surface temperature

Action

BK Medical notified customers by letter on 4/10/06 via FedEx advising user of the problem and scheduling the software correction by a Service representative visit .

Distribution

CA,CT, FL, MD, ME, MN, MS, NY, OH, RI, SD, TX, VA

Quantity

28 units