Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667
Recall: Z-0856-06
·
Initiated March 16, 2006
Recall
- Recall Number
- Z-0856-06
- Event Number
- 34900
- Firm
- B & K Medical Systems, Inc.
- FEI Number
- 1000305740
- Product Code
- IYN
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 16, 2006
- Posted
- May 9, 2006
- Terminated
- October 12, 2006
- Address
- 250 Andover St, Wilmington, MA, 01887-1048
Description
Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667
Reason
Due to a problem with calibration, the device can produce up to 50% elevated acoustic output and up to 2 degrees higher surface temperature
Action
BK Medical notified customers by letter on 4/10/06 via FedEx advising user of the problem and scheduling the software correction by a Service representative visit .
Distribution
CA,CT, FL, MD, ME, MN, MS, NY, OH, RI, SD, TX, VA
Quantity
28 units