FDA Recall Terminated

Eclipse 8.5 (build 8.2.2X) systems with distributed Calculation Framework (DCF) 8.2.22, model number H48, manufactured by Varian Medical Systems, Inc., Palo Alto, CA. Indicated for planning proton treatment of neoplasms of the eye.

Recall: Z-0855-2009 · Initiated October 2, 2008

Recall

Recall Number
Z-0855-2009
Event Number
50143
Firm
Varian Medical Systems Oncology Systems
FEI Number
2916710
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
October 2, 2008
Posted
January 15, 2009
Terminated
November 27, 2010
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

Eclipse 8.5 (build 8.2.2X) systems with distributed Calculation Framework (DCF) 8.2.22, model number H48, manufactured by Varian Medical Systems, Inc., Palo Alto, CA. Indicated for planning proton treatment of neoplasms of the eye.

Reason

When using the multiple static segment option, the device may alter dose delivery.

Action

An Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps on October 2, 2008. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. The firm recommends that customers DO NOT USE the Multiple Static Segment (MSS) option in Varian Leaf Motion Calculator version 8.2.22. The firm also recommends setting the default delivery method for Varian Leaf Motion Calculator to Sliding Window. When using Eclipse 8.5, always follow the Safety Instructions found in chapter 2 of "Eclipse External Beam Planning Instruction for Use", P/N B501357R01A, March 2008. If customers believe a patient treatment has been affected by this issue, they should notify a Varian Service or Support person as soon as possible.

Distribution

Worldwide Distribution --- including USA and countries of Switzerland and the Netherlands.

Quantity

64 units