FDA Recall Terminated

Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2)

Recall: Z-0854-06 · Initiated October 8, 2004

Recall

Recall Number
Z-0854-06
Event Number
34516
Firm
General Electric Med Systems LLC
FEI Number
2126677
Product Code
LLZ
Status
Terminated
Root Cause
Other
Initiated
October 8, 2004
Posted
May 9, 2006
Terminated
December 31, 2006
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2)

Reason

Display in fused mode- viewer allows 2 studies ( PET/CT) registered in the same spatial domain to be viewed simultaneously. These 2 images should be visually aligned to a precision of less than 1/2 PET Voxel. A problem associated with this version of software may introduce a shift of up to 1.5 PET Voxels, which can be as much as 8 mm with certain zoom factors.

Action

An Urgent Safety Notice letter, dated October 8, 2004,(5121309-147, rev.E) was distributed to affected customers and a validated software patch that corrected the problem was installed on affected systems via Field Modification Instruction (FMI) 80114.

Distribution

Distributed in 32 states: AK, AL, AZ, CA, CO, CT, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, NC, ND, NH, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV; Australia, Belgium, Brazil, Canada, Switzerland, Chile, Nigeria, China, France, Egypt, Ireland, Israel, Japan, Korea, the Netherlands, Norway, New Zealand, Phillipines, Sweden, and Taiwan.

Quantity

199