FDA Recall
Terminated
STA Liatest VWF:Ag, Product Catalog Number : 0518. This is a Immuno-Turbidimetric Assay of von Willebrand Factor.
Recall: Z-0853-2007
·
Initiated March 13, 2007
Recall
- Recall Number
- Z-0853-2007
- Event Number
- 37746
- Firm
- Diagnostica Stago, Inc.
- FEI Number
- 2245451
- Product Code
- GGP
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 13, 2007
- Posted
- July 3, 2007
- Terminated
- January 16, 2009
- Address
- 5 Century Dr, Parsippany, NJ, 07054-4607
Description
STA Liatest VWF:Ag, Product Catalog Number : 0518. This is a Immuno-Turbidimetric Assay of von Willebrand Factor.
Reason
A dose-hook effect is observed starting from VWF:AG levels of 300%, while the package insert indicates that no dose-hook effect has been observed with VWF:Ag levels up to 800%.
Action
Customers were notified on March 13, 2007 by letter, first class mail.
Distribution
Hospitals and laboratories Nationwide
Quantity
1,098 units