Affymetrix GeneChip Microarray Instrumentation System, consisting of GeneChip 3000Dx scanner with autoloader, FS450Dx fluidics station and GCOSDX Software

Recall: Z-0853-06 · Initiated April 26, 2005

Recall

Recall Number
Z-0853-06
Event Number
35175
Firm
Affymetrix, Inc.
FEI Number
3003314809
Product Code
NSU
Status
Terminated
Root Cause
Other
Initiated
April 26, 2005
Posted
May 6, 2006
Terminated
December 13, 2006
Address
4G Crosby Dr, Bedford, MA, 01730-1402

Description

Affymetrix GeneChip Microarray Instrumentation System, consisting of GeneChip 3000Dx scanner with autoloader, FS450Dx fluidics station and GCOSDX Software

Reason

Incorrect software version of instrument controller is not compatible with new configuration. This incorrect configuration may cause the instrument system to fail at start-up or during a run.

Action

Afffymetrix notified the two hospitals by email dated 4/14/05 and by letter to the Distributor on 4/28/05. Affymetrix service personnel will be provided with discs to perform the software update. All field changes will be documented via Service Work Orders.

Distribution

California and Indiana Foreign: The Netherlands, Switzerland

Quantity

13 units