FDA Recall Open, Classified

Hanger-Bar 2P 45cm

Recall: Z-0851-2022 · Initiated November 1, 2021

Recall

Recall Number
Z-0851-2022
Event Number
89509
Firm
Human Care USA, Inc.
FEI Number
3002833166
Product Code
FSA
Status
Open, Classified
Root Cause
Process control
Initiated
November 1, 2021
Address
4210 S Industrial Dr, Ste 160, Austin, TX, 78744-1077

Description

Hanger-Bar 2P 45cm

Reason

The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.

Action

On 11/01/2021, the firm emailed an "URGENT PRODUCT CORRECTION" Notification to customers/distributors informing them that Loctite may be missing on the stop-screw which holds the hanger-bar hook in place. This is the hook which connects to the distance strap. Because there is a possibility of the screw coming loose from the hook in that case the user may fall. If the user falls there is a risk of minor to more serious injury. This depends on the fall level, the surface user falls on and the strength and state of health of the user. Customers are instructed to: 1. Unscrew the M4 screw (fig. 1) and check the guide threads (see yellow arrow in fig. 2) and that Loctite has been applied to the M4 screw. If the M4 screw is missing or the threads are of poor quality (worn or damaged), the product should not be used. Contact the Recalling Firm. 2. Apply Loctite 243 (fig. 3) on the M4 screw (fig. 4). 3. Screw on the M4 screw again with a torque of 1.5 Nm (fig. 5) until it stops. Wait a few minutes until Loctite has dried. Then check that the hook can be swung freely (fig. 6). 4. The Recalling Firm requests that all customers confirm that they have received the Recall Notice and that the necessary action has been taken. A response form was sent along with Recall Notice. Send completed response forms to [email protected] For any questions, call - 512-476-7199 ext. 290

Distribution

U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV.

Quantity

10 devices