FDA Recall
Terminated
GE Single 15 inch TV Monitor Suspension for XT Bridge, Model 46-240485P2, Catalog Number B2056HA
Recall: Z-0851-04
·
Initiated March 30, 2004
Recall
- Recall Number
- Z-0851-04
- Event Number
- 28753
- Firm
- General Electric Med Systems
- FEI Number
- 2126677
- Product Code
- EAM
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 30, 2004
- Posted
- July 20, 2004
- Terminated
- July 24, 2006
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Single 15 inch TV Monitor Suspension for XT Bridge, Model 46-240485P2, Catalog Number B2056HA
Reason
The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.
Action
Field modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals.
Distribution
The units were distributed nationwide in the United States and worldwide.
Quantity
1236 devices