FDA Recall Terminated

Kaguya Automated Peritoneal Dialysis System Product Usage: It is intended for automatic control of dialysate solution exchanges in the treatment of adult renal failure patients undergoing Peritoneal dialysis.

Recall: Z-0850-2019 · Initiated December 11, 2018

Recall

Recall Number
Z-0850-2019
Event Number
81738
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
FKX
Status
Terminated
Root Cause
Software design
Initiated
December 11, 2018
Terminated
February 12, 2021
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

Kaguya Automated Peritoneal Dialysis System Product Usage: It is intended for automatic control of dialysate solution exchanges in the treatment of adult renal failure patients undergoing Peritoneal dialysis.

Reason

Potential for the software on Automated PD System cyclers which can cause shortened dwell times during a Cycle-Based therapy

Action

Baxter Healthcare Corporation notified providers on about 12/11/2018 of the recall concerning the AMIA Automated Peritoneal Dialysis System, via "Urgent Medical Device Correction" letter. Patients were notified on about 1/14/2018 via "Urgent Medical Device Correction" letter. Providers were instructed to continue using the AMIA Automated PD System however, for patients who are prescribed a Cycle-Based therapy, to change these therapies to Time-Based by following the programming instructions within the Clinician Guide, Section 5.3.1 - Time-Based Programming Method. In Time-Based therapy, to maintain a fixed dwell time, the Smart Dwell treatment option needs to be Disabled. To disable the Smart Dwell treatment option, follow the programming instructions within the Clinician Guide, Section 5.3.1 - Time-Based Programming Method. Questions can be directed to Baxter's Renal Clinical Helpline at 888-736-2543, option 2, Monday through Friday, 8:00 am to 4:30 pm (CT). Providers were informed that Baxter would contact them when the software update is available and provide further instructions on how to download the upgrade remotely. They were also instructed to complete and return the Baxter Customer Reply Form and to inform customers if the device was further distributed to other facilities or departments within their institution. Patients were instructed to continue using the AMIA Automated PD System however, they were directed to check the Programming Method to ensure they have not been prescribed Cycle-Based therapy. The programming method can be verified by following section 10.3 of the Patient Guide: 1. touch "GO" on the home screen and the Confirm Treatment screen should appear, 2. touch "Program Settings" to view the program settings details and the Program Details screen will appear, 3. display the "Program Method" by navigating to Page 3 by pressing the down arrow at the bottom of the screen. If the Programming Method is Time-Based or Tidal, the iss

Distribution

Worldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to Canada and Japan.

Quantity

1,085 devices