FDA Recall Terminated

Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform.

Recall: Z-0850-2015 · Initiated November 6, 2014

Recall

Recall Number
Z-0850-2015
Event Number
69715
Firm
Illumina Inc
FEI Number
3003218906
Product Code
PFS
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
November 6, 2014
Posted
December 23, 2014
Terminated
March 17, 2015
Address
5200 Illumina Way, San Diego, CA, 92121

Description

Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform.

Reason

Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument.

Action

A customer notification letter dated 11/12/14 was sent to all customers who purchased the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay. The letter confirms that there is a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument. At certain genomic locations within the region sequenced by the assay, deletions cannot be reported by the software. The letter outlines the potential issues, the results of Illuminia's investigation, and the firm's next steps. Customers' with technical questions or concerns, are instructed to contact their local support team or Illumina Technical Support at [email protected]. Customers with questions regarding the field action, are instructed to contact Illumina Regulatory Affairs at [email protected]. Customers will be contacted by 11/13/14 by phone and a phone script was provided that includes the procedure, talking points, actions required, and next steps.

Distribution

US: MD, NY Non-US: France

Quantity

19 units