FDA Recall Terminated

Sodasorb Carbon Dioxide Absorbent, Medical Grade (16%) Soda Lime USP-NF with an ethyl violet indicator which turns from white to violet as absorber is depleted; Manufactured by: W.R. Grace & Co. - Conn., 6050 West 51st St., Chicago, IL 60638; the product is packaged in pre-paks(pre-filled 2.5 lb. canisters, 12 per carton) and 5-gallon pails

Recall: Z-0850-05 · Initiated March 31, 2005

Recall

Recall Number
Z-0850-05
Event Number
31687
Firm
Darex Container Products Div of W.R. Grace & Co.
FEI Number
3003940942
Product Code
CBL
Status
Terminated
Root Cause
Other
Initiated
March 31, 2005
Posted
May 25, 2005
Terminated
July 31, 2006
Address
6050 W 51st St, Chicago, IL, 60638-1405

Description

Sodasorb Carbon Dioxide Absorbent, Medical Grade (16%) Soda Lime USP-NF with an ethyl violet indicator which turns from white to violet as absorber is depleted; Manufactured by: W.R. Grace & Co. - Conn., 6050 West 51st St., Chicago, IL 60638; the product is packaged in pre-paks(pre-filled 2.5 lb. canisters, 12 per carton) and 5-gallon pails

Reason

Both lots were found to be out of specification due to an elevated level of small particles (fines).

Action

Recall letters dated 3/30/05 were sent to the accounts via DHL on 3/31/05, informing them that the two lots did not undergo final QC testing prior to shipment and do not meet finished product specifications. The accounts were requested to check their inventory for the affected lots, discontinue use and place it on hold pending return to W.R. Grace for replacement. Any questions were directed to the Sodasorb Recall Coordinator at 1-877-327-3987.

Distribution

California, Connecticut, Pennsylvania, Maine, Massachusetts, MIssissippi, New Hampshire, New Jersey, New York, Vermont, Virginia and West Virginia

Quantity

360 PrePaks and 18 5-gallon pails