FDA Recall Open, Classified

Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.

Recall: Z-0848-2026 · Initiated November 3, 2025

Recall

Recall Number
Z-0848-2026
Event Number
97939
Firm
Fresenius Kabi USA, LLC
FEI Number
3014732157
Product Code
FRN
Status
Open, Classified
Root Cause
Labeling design
Initiated
November 3, 2025
Posted
December 1, 2025
Address
50 High St, Ste 50, North Andover, MA, 01845-2620

Description

Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.

Reason

Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.

Action

Fresenius Kabi notified consignees on about 11/03/2025 via emailed letter. Consignees were informed of the issue and instructed to account for additional backpressure if check valve is used when configuring the downstream occlusion pressure alarm setting, adjust the occlusion alarm pressure to allow for a more balanced approach between occlusion detection and avoidance of unnecessary alarms, and continue to follow all Instructions for Use and institutional policies regarding infusion pump setup, monitoring, and alarm response. Consignees were requested to ensure all relevant persons within the organization are informed about this letter and the actions as described per the enclosed Tip Sheet on How to Use a Check Valve with the Ivenix Pump & Administration Sets. Additionally, consignees were requested complete and return the Customer Reply Form. Fresenius Kabi plans to update the Instructions for Use to include the information presented in the "Required Actions for Users", "Important Note on Check Valve Usage" and/or Tip Sheet on "How to Use a Check Valve with the Ivenix Pump & Administration Sets" for which scheduling is still being finalized.

Distribution

US Nationwide distribution in the sates of CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, UT, VA, WA, WI.

Quantity

15,862 units