Cardinal Health Monoject, 6ML Enteral Syringe (purple) Low Dose (LDT) - For Use with ENFit Connection System, REF 406SE
Recall
- Recall Number
- Z-0847-2024
- Event Number
- 93676
- Firm
- Cardinal Health 200, LLC
- FEI Number
- 3001236905
- Product Code
- PNR
- Status
- Open, Classified
- Root Cause
- Process change control
- Initiated
- December 28, 2023
- Posted
- February 2, 2024
- Address
- 3651 Birchwood Dr, Waukegan, IL, 60085-8337
Description
Cardinal Health Monoject, 6ML Enteral Syringe (purple) Low Dose (LDT) - For Use with ENFit Connection System, REF 406SE
The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal Health recommends that they not be used with enteral syringe feeding pumps.
Cardinal Health issued an URGENT MEDICAL DEVICE PRODUCT CORRECTION notice to its consignees on 12/28/2023 via FedEx. The notice explained the problem with the device, risk to patient, and requested the following actions be taken: 1. REVIEW your inventory for the affected product codes and lots. Location of product code and lot are shown in above listed table and below example labeling. 2. COMMUNICATE with all personnel that utilize the Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) that they should not be used with enteral syringe feeding pumps. 3. POST a copy of this notification in your storeroom and clinical areas where the product is stored. 4. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this medical device product correction and share a copy of this notice. 5. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected], whether you have affected product or not. CONTACT the appropriate Customer Service group with questions related to this notification. Monday Friday between 8:00am - 5pm EST: " Hospital 800-965-5227 " Distributor 800-635-6021 " All Other Customers 888-444-5440 For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332. A subsequent recall letter dated 2/2/2024 flagged URGENT MEDICAL DEVICE PRODUCT RECALL was issued informing the customer that the recall now involves all sizes of the Cardinal Health brand Monoject sterile Syringe Luer-Lock Soft Packs and Monoject sterile Enteral Syringes with the ENFit connection. This letter supersedes the previous communications. The letter explains the issue, risk to health, and lists the actions that are required. The actions were: 1. REVIEW your inventory for the
Distribution US nationwide and Canada. Foreign distribution updated 9/6/2024 to only include Canada.
20186 units