FDA Recall Terminated

VITROS XT Chemistry Products ALB-TP Slides (ALB is not affected) - IVD quantitatively measures total protein (TP) concentration in serum and plasma using VITROS XT 3400 Chemistry Systems and VITROS XT 7600 Integrated Systems Product Code: 6844293

Recall: Z-0845-2021 · Initiated December 15, 2020

Recall

Recall Number
Z-0845-2021
Event Number
87019
Firm
Ortho Clinical Diagnostics Inc
FEI Number
2250051
Product Code
CEK
Status
Terminated
Root Cause
Device Design
Initiated
December 15, 2020
Terminated
July 27, 2021
Address
1001 US Highway 202, Raritan, NJ, 08869-1424

Description

VITROS XT Chemistry Products ALB-TP Slides (ALB is not affected) - IVD quantitatively measures total protein (TP) concentration in serum and plasma using VITROS XT 3400 Chemistry Systems and VITROS XT 7600 Integrated Systems Product Code: 6844293

Reason

Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides

Action

Ortho-Clinical Diagnostics (Ortho) issued customer letter (Ref. CL2018-308) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) on 15 Dec 2020 to all customers who received one of the affected products in the previous 18 months. Letter states reason for recall, health risk and action to take: Complete the enclosed Confirmation of Receipt form no later than insert date. Please provide the enclosed communication to customers who were shipped the affected products (listed above) from your facility within the last 18 months. Ortho has communicated with your customers that received drop shop product from our warehouse. Foreign affiliates were informed by email on 15 Dec 2020 of the issue and instructed to notify their consignees of the issue and their required action until the updated the IFU is available.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of AR, AU, BM, BR, CA, CL, CN, CO, CR, DO, EC, FR, FR, GT, HT, IN, JP, MX, PA, PE, PR, PY, SV.

Quantity

US=9802; OUS: 1298 Total: 11,100