Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
Recall
- Recall Number
- Z-0845-2019
- Event Number
- 81950
- Firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- FEI Number
- 2182208
- Product Code
- OSR
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- January 17, 2019
- Terminated
- March 30, 2021
- Address
- 8200 Coral Sea St Ne, Mounds View, MN, 55112-4391
Description
Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.
In the US, beginning 17-Jan-2019, Medtronic Field Representatives (Field Reps) will locate all Encore programmers both in their trunk stock and in consignment locations. Since the Encore Power Supplies are interchangeable among all Encores, all Encore programmer distributed units are being located, and the associated Power Supplies will be checked to identify any affected units. If the Field Rep identifies an affected Power Supply physically at a customer account, the Medtronic Field Rep will hand deliver a courtesy FCA Notification Letter with instructions to the Risk Manager or HealthCare Professional at the account.
US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic
731 units