FDA Recall Terminated

Computed Tomography X-Ray Systems (Brilliance 64, Model Number: 728231; Brilliance Big Bore, Model Number: 728243; Brilliance iCT, Model Number: 728306; Brilliance iCT SP, Model Number: 728311; and Ingenuity CT, Model Number: 728326); Mfr. by Philips Medical Systems; Cleveland, OH. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Recall: Z-0845-2012 · Initiated October 7, 2011

Recall

Recall Number
Z-0845-2012
Event Number
60873
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
October 7, 2011
Posted
January 24, 2012
Terminated
June 18, 2013
Address
595 Miner Road, Cleveland, OH, 44143-2131

Description

Computed Tomography X-Ray Systems (Brilliance 64, Model Number: 728231; Brilliance Big Bore, Model Number: 728243; Brilliance iCT, Model Number: 728306; Brilliance iCT SP, Model Number: 728311; and Ingenuity CT, Model Number: 728326); Mfr. by Philips Medical Systems; Cleveland, OH. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason

The Common Image Reconstruction System (CIRS) on the affected Brilliance Systems encounters an error and requires reboot of the system. The unit failed to initialize axial scan due to timeout that occurred waiting for "READY" state

Action

On 10/10/2011 the firm sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected products, the problem, the hazard involved, the action to be taken by the customer, and actions planned by Philips. Customers were to reboot their systems if the two possible scenarios occurred. Philips Healthcare plans on installing the update through a Field Change Order free of charge. For further information or support concerning the issue, contact a local Philips representative or local Philips Healthcare office. In North America and Canada, contact Customer Care Solutions Center at 1-800-722-9377, option 5.

Distribution

Worldwide Distribution -- USA, including the states of: AL, AZ, CA, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA & WV and the countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BANGLADESH, BELGIUM, Brazil, CHINA, COLUMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KOREA, LUXEMBOURG, MALAYSIA, MALI, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PHILIPPINES, POLAND, QATAR, RUSSIA, SINGAPORE, SLOVENIA, SOUTH KOREA, SPAIN, SUDAN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN & VIETNAM.

Quantity

548 units