Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT
Recall
- Recall Number
- Z-0844-2024
- Event Number
- 93768
- FEI Number
- 1282497
- Product Code
- GAR
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- December 21, 2023
- Posted
- January 29, 2024
- Address
- LP 15 Hampshire St, Covidien, Mansfield, MA, 02048-1113
Description
Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
On December 21, the firm notified affected consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to identify, quarantine, and return all unused, non-expired affected sutures to Medtronic. Please contact [email protected] for the Return Good Authorization (RGA). Customers will receive credit for returned affected product. If purchased from a distributor, please contact your distributor directly to arrange for the return of product back to your distributor. Pass on the recall notice to all those who need to be aware within your organization, or to any organization where the affected sutures have been transferred or distributed. If you have questions regarding this recall, please call Medtronic Quality Assurance at 800-255-6774 option 1, then option 1 ***Updated 4/1/2024*** Firm identified additional models and lots affected by the gamma radiation issue and sent additional notification to customers on January 25, 2024.
Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
46,812 US; 68,976 OUS