Luer Male Large Bore, REF/CAT No.: 102100001, for use in fluid administration sets, 40000.
Recall: Z-0844-2007
·
Initiated April 17, 2007
Recall
- Recall Number
- Z-0844-2007
- Event Number
- 37902
- Firm
- Merit Medical Systems, Inc.
- FEI Number
- 1721504
- Product Code
- FMG
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 17, 2007
- Posted
- May 31, 2007
- Terminated
- May 5, 2008
- Address
- 1600 Merit Pkwy South, Jordan, UT, 84095-2416
Description
Luer Male Large Bore, REF/CAT No.: 102100001, for use in fluid administration sets, 40000.
Reason
Male luer in fluid administration sets may crack due to inadequate molding process
Action
Consignees were notified by letter on 04/17/2007 and instructed to quarantine and return any unused product.
Distribution
CA, IL, NJ, NY. No military, government or foreign distribution.
Quantity
40,000 units