Suction Instrument with Tubing, Catalog No. 303. Instrument: Yankauer with Vent. *Tubing: 6'' x 1/4'' I.D. Non-Conductive. Sterile, Single Use, Latex Free. Rx ONLY. *Tubing Made in Mexico. busse Hospital Disposables Hauppauge, NY 11788. --- Classification Name: Unit, Suction Operatory. Class I device, 510 (k) exempt, Device Listing Number A715729. Product comprises 2 components: a Yankauer made from K-Resin and tubing made from PVC. Each instrument is packed in a flexible packaging comprising Tyvek(R) and plastic film. There are 20 units per shipping carton.
Recall
- Recall Number
- Z-0844-04
- Event Number
- 28815
- Firm
- Busse Hospital Disposables
- FEI Number
- 2433012
- Product Code
- EBR
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 13, 2004
- Posted
- July 20, 2004
- Terminated
- February 18, 2005
- Address
- 75 Arkay Drive, Hauppauge, NY, 11788-3707
Description
Suction Instrument with Tubing, Catalog No. 303. Instrument: Yankauer with Vent. *Tubing: 6'' x 1/4'' I.D. Non-Conductive. Sterile, Single Use, Latex Free. Rx ONLY. *Tubing Made in Mexico. busse Hospital Disposables Hauppauge, NY 11788. --- Classification Name: Unit, Suction Operatory. Class I device, 510 (k) exempt, Device Listing Number A715729. Product comprises 2 components: a Yankauer made from K-Resin and tubing made from PVC. Each instrument is packed in a flexible packaging comprising Tyvek(R) and plastic film. There are 20 units per shipping carton.
Firm received complaints of "melting" at tubing connection. The connecting part of the Yankauer becomes soft after extended direct contact with the PVC tubing. This results in a slower rate of suction.
Urgent Product Recall Letters and Response Forms were sent to the consignees via Registered Mail on 4/13/04.
Distributors and end users (hospitals, fire departments/EMS) located nationwide, in Puerto Rico, St. Thomas, Canada, Dominican Republic, Panama, Uganda, Dubai, and Abu Dhabi, United Arab Emirates.