AEROSET Analyzer

Recall: Z-0844-03 · Initiated August 15, 2002

Recall

Recall Number
Z-0844-03
Event Number
25771
Firm
Abbott Laboratories, Inc
FEI Number
1628664
Product Code
JJE
Status
Terminated
Root Cause
Other
Initiated
August 15, 2002
Posted
May 20, 2003
Terminated
April 7, 2004
Address
1921 Hurd Drive PO Box 2020, Irving, TX, 75038

Description

AEROSET Analyzer

Reason

Design of high concentration waste line causes formation of a precipitate in the waste line causing overflow of waste.

Action

The firm sent Technical Service Bulletins dated August 15, 2002 to consignees. Field Representatives are correcting the device in the field. A customer letter was released on August 20, 2002 to provide customers actions to take to prevent the occurrance.

Distribution

Nationwide and to the following foreign countries: Mexico, Venezuela, Columbia, Argentina, Brazil, Puerto Rico, Dominican Republic, Canada, Germany, Hong Kong, New Zealand, and China

Quantity

578