FDA Recall
Terminated
AEROSET Analyzer
Recall: Z-0844-03
·
Initiated August 15, 2002
Recall
- Recall Number
- Z-0844-03
- Event Number
- 25771
- Firm
- Abbott Laboratories, Inc
- FEI Number
- 1628664
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 15, 2002
- Posted
- May 20, 2003
- Terminated
- April 7, 2004
- Address
- 1921 Hurd Drive PO Box 2020, Irving, TX, 75038
Description
AEROSET Analyzer
Reason
Design of high concentration waste line causes formation of a precipitate in the waste line causing overflow of waste.
Action
The firm sent Technical Service Bulletins dated August 15, 2002 to consignees. Field Representatives are correcting the device in the field. A customer letter was released on August 20, 2002 to provide customers actions to take to prevent the occurrance.
Distribution
Nationwide and to the following foreign countries: Mexico, Venezuela, Columbia, Argentina, Brazil, Puerto Rico, Dominican Republic, Canada, Germany, Hong Kong, New Zealand, and China
Quantity
578